Guidance and manufacturer's declaration — Electromagnetic
immunity
This model is intended for use in the electromagnetic environment specified below. The
customer or the user of this model should assure that it is used in such an environment.
IEC 60601
Immunity test
test level
Electrostatic
Contact:
2, 4, 6 kV
discharge (ESD)
IEC 61000-4-2
Air:
2, 4, 8 kV
2 kV
Electrical fast
transient/burst
for power supply lines
IEC 61000-4-4
1 kV
for input/output lines
Surge
Differential mode:
0.5, 1 kV
IEC 61000-4-5
Common mode:
0.5, 1, 2 kV
Voltage dips, short
< 5% U
T
interruptions, and
(> 95% dip in U
voltage variations
for 0.5 cycle
on power supply
40% U
T
input lines
(60% dip in U
IEC 61000-4-11
for 5 cycle
70% U
T
(30% dip in U
for 25 cycle
< 5% U
T
(> 95% dip in U
for 5 seconds
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
Definition
U
is the a.c. mains voltage prior to application of the test level.
T
BF-190 Series OPERATION MANUAL
Compliance level
Same as left
Same as left
Same as left
Same as left
)
T
)
T
)
T
)
T
Same as left
EMC information
Electromagnetic environment —
Guidance
Floors should be made of wood, concrete, or
ceramic tile that hardly produces static. If
floors are covered with synthetic material that
tends to produce static, the relative humidity
should be at least 30%.
Mains power quality should be that of a typical
commercial (original condition feeding the
facilities) or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the
user of this instrument requires continued
operation during power mains interruptions, it
is recommended that this instrument be
powered from an uninterruptible power supply
or a battery.
It is recommended to use this instrument by
maintaining enough distance from any
equipment that operates with high current.