Summary of Contents for Sanwa RD700

  • Page 1 RD700 RD701 DIGITAL MULTIMETER INSTRUCTION MANUAL...
  • Page 2: Table Of Contents

    CONTENS SAFETY PRECAUTIONS *Before use, read the following safety precautions. 1-1 Explanation of Warning Symbols ......1 1-2 Warning Instruction for Safe Use.
  • Page 3 [8] SPECIFICATIONS General Specification ........19 Measurement Range and Accuracy .
  • Page 4: Safety Precautions

    *Before use, read the following safety precautions. This instruction manual explains how to use your new digital multimeter RD700 or RD701 safely. Before use, please read this manual thoroughly. After reading it, keep it together with the product for reference to it when necessary.
  • Page 5: Overload Protections

    WARNING 9. Always keep your fingers behind the finger guards on the probe when making measurements. 10. Be sure to disconnect the test pins from the circuit when changing the function. 11. Before starting measurement, make sure that the function and range are properly set in accordance with the measurement.
  • Page 6: Name Of Component Units

    [ 3 ] NAME OF COMPONENT UNITS LCD Display RELATIVE Button SELECT Button DATA Hold Button / MAX Hold Button RANGE Hold Button POWER Switch Holster and FUNCTION Switch V Hz ADP A Measuring TEMP Measuring Terminal Terminal µA mA Measuring Terminal Common Input Terminal Test Pin Test Probe (RED)
  • Page 7: Description Of Functions

    [ 4 ] DESCRIPTION OF FUNCTIONS 4-1 Function Switch Turn this switch to turn on and off the power and to select the functions of "V / V / Hz / ADP / / TEMP / µA / mA / A". 4-2 Auto Power Off The meter will enter a low power consumption sleep mode automatically to extend battery life after approximately 30 minutes of no function switch or...
  • Page 8: Relative Mode: Rel

    (like voltage measurement function) is selected. 4-10 Words AC Sensing Average: RD700 When measurement is taken by "average", no error is caused as the input signal is shine wave with no distortion. However, if the input waveform is distorted sine cave or non-sinusoidal wave, conversion to root-mean -square values is very difficult, resulting in a large error.
  • Page 9: Measurement Procedure

    NMRR (Normal Mode Rejection Ratio) NMRR is the DMM's ability to reject unwanted AC noise effect, which can cause inaccurate DC measurements. NMRR is typically specified in terms of dB (decibel). This series has a NMRR specification of >60dB at 50 and 60Hz, which means a good ability to reject the effect of AC noise in DC measurements.
  • Page 10: Voltage Measurement / Frequency Measurement

    5-2 Voltage measurement / Frequency measurement WARNING 1. Never apply an input signal exceeding the maximum rating input value. 2. Be sure to disconnect the test pins from the circuit when changing the function. 3. Always keep your fingers behind the finger guards on the probe when making measurements.
  • Page 11 [DCV Measurement] [ACV Measurement] 5-2-2 Frequency measurement 1) Application Frequency of an AC circuit is measured. 2) Measuring ranges 10.00Hz to 1.000MHz (Auto range) 3) Measurement procedure Connect the plug of black test lead to COM measuring input terminal and plug of red test lead to Hz measuring terminal. Set the function switch at 'Hz'.
  • Page 12: Adapter Measurement

    5-3 Adapter measurement 1) Applications This function is to measure with optional clamp probe (CL-22AD and CL33DC), and like. It can measure with this function, if it is the probe like the following. Current clamp probe with 1mV (Up to 400mV) output per 1A. 2) How to use Connect the plug of black test lead or '-' to COM input terminal and plug of red test lead or '+' to ADP input terminal.
  • Page 13 [Adapter (ADP) indication example] Clamp Probe Range Probe Output DMM Display Reading value DC 20A* DC 15mV DC 0150 1.5A DC 200A DC 150mV DC 1500 150A CL-22AD AC 20A* AC 15mV AC 0150 1.5A AC 200A AC 150mV AC 1500 150A DC 30A* DC 25mV...
  • Page 14: Resistance Measurement And Capacitance Measurement

    5-4 Resistance measurement and Capacitance measurement / Testing Diode / Checking Continuity WARNING Never apply voltage to the input terminals. 5-4-1 Resistance Measurement 1) Applications Resistance of resistors and circuits are measured. 2) Measuring ranges 6 ranges from 400.0 to 40.00M . 3) Measurement procedure Connect the plug of black test lead to COM input terminal and plug of red test lead to...
  • Page 15 5-4-2 Checking Continuity 1) Application Checking the continuity of wiring and selecting wires. 2) How to use Connect the plug of black test lead to COM measuring input terminal and plug of red test lead to measuring terminal. Set the function switch at ' / / ' and select ' ' with the SELECT button.
  • Page 16 5-4-3 Testing Diode 1) Application The quality of diodes is tested. 2) How to use Connect the plug of black test lead to COM measuring input terminal and plug of red test lead to measuring terminal. Set the function switch at ' / / ' and select ' ' with the SELECT button.
  • Page 17 5-4-4 Capacitance Measurement CAUTION 1. Discharge the capacitance before measurement. 2. Due to the measurement method that this device pours the current into a capacitor the measurement such as the big electrolysis capacitor of a leakage current is not suited in order that the error becomes big.
  • Page 18: Temperature Measurement

    5-5 Temperature Measurement 1) Application The temperature of outside and water, object is measured. 2) Measuring ranges Centigrade: Range from -20 C to 300 C Fahrenheit: Range from -4 F to 572 F 3) Measurement procedure Input the -plug to COM input terminal and the +plug to TEMP terminal.
  • Page 19: Current Measurement

    5-6 Current Measurement WARNING 1. Never apply voltage to the input terminals. 2. Be sure to make a series connection via load. × 3. Do not apply an input exceeding the maximum rated current to the input terminals. 4. Before starting measurement, turn OFF the power switch of the circuit to separate the measuring part and connect the test leads firmly.
  • Page 20: Maintenance And Inspection

    For measurement of DCA, apply the black test pin to the negative potential side of the circuit to measure and the red test pin to the positive potential side in series with load. For measurement of ACA, apply the red and black test pins to the circuit to measure in series with load.
  • Page 21: Calibration

    6-2 Calibration The manufacturer may conduct the calibration and inspection. For more information, please contact the manufacturer. 6-3 Battery and Fuse Replacement WARNING 1. If the rear case or the battery lid is removed with input applied to the input terminals, you may get electrical shock. Before starting the work, always make sure that no input is applied.
  • Page 22: After-Sale Service

    (specified accuracy)/ C @ (0 C or 28 C~50 C), or otherwise specified Power Supply: 9V battery; NEDA1604, JIS006P or IEC6F22 Sensing: Average Sensing for RD700. True RMS for RD701 Safety: Meets IEC61010-1 (1995), EN61010-1 (1995), to terminals: - 19 -...
  • Page 23 Total Accuracy = Specified Accuracy + 40 digits Performance above 3V/m is not specified Power Consumption: 3.2mA Typical Auto Power Off Timing: Idle for 30 minutes Auto Power Off Function Consumption: 300µA typical for RD700; 360µA typical for RD701 Dimension: 179(L) 87(W)
  • Page 24: Measurement Range And Accuracy

    8-2 Measurement Range and Accuracy Accuracy assurance range: 23±5 & less than 75% R.H. Input Function & Range Accuracy Remarks Impedance 400.0mV ±(0.3%rdg + 4dgt) 1000M NMRR: 4.000V >50dB@50/60Hz 10M , ±(0.5%rdg + 3dgt) 40.00V CMRR: >120dB@DC, 30pF 400.0V 50/60Hz, Rs=1k nominal 1000V ±(1.0%rdg + 4dgt)
  • Page 25 Input Function Accuracy Remarks Impedance ±(0.3%rdg + 4dgt) 1000M , 30pF RD700: 50Hz~500Hz nominal ±(1.8%rdg + 5dgt) RD701: 50Hz~3kHz 3) The accuracy of the sensor is not included. Accuracy calculation For example… Measurement 400mVDC Range Reading value : 100.0 [mV] Accuracy : 400.0[mV] Range…±(0.3%rdg + 4dgt)
  • Page 26 SANWA ELECTRIC INSTRUMENT CO.,LTD. Dempa Bldg,Sotokanda2-Chome Chiyoda-Ku,Tokyo,Japan 02.04...