Electromagnetic Immunity Declaration (Eid - ZOLL SurePower Operator's Manual

Charger station
Hide thumbs Also See for SurePower:
Table of Contents

Advertisement

M
'
D
ANUFACTURER
S
ECLARATION

Electromagnetic Immunity Declaration (EID)

Immunity Test
Electrostatic
discharge (ESD)
IEC 61000 4-2
Electrical fast
transient/burst
IEC 61000 4-4
Surge
IEC 61000 4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
U
is the AC mains voltage prior to application of the test level.
T
A-2
-- E
E
LECTROMAGNETIC
IEC 60601 Test
Level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 1 kV differential
mode
± 2 kV for common
mode
<5% U
(>95% dip
T
in U
) for 0.5 cycle
T
40% U
(60% dip in
T
U
) for 5 cycles
T
70% U
(30% dip in
T
U
) for 25 cycles
T
<5% U
(>95% dip
T
in U
) for 5 sec.
T
3 A/m
www.zoll.com
I
MISSIONS AND
MMUNITY
Compliance Level
Electromagnetic
Environment - Guidance
± 6 kV contact
Floors should be wood,
± 8 kV air
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
± 2 kV for power
Mains power quality
supply lines
should be that of a typical
commercial or hospital
± 1 kV
environment.
± 1 kV differential
Mains power quality
mode
should be that of a typical
commercial or hospital
± 2 kV for common
environment.
mode
<5% U
(>95% dip
Mains power quality
T
should be that of a typical
in U
) for 0.5 cycle
T
commercial or hospital
40% U
(60% dip in
T
environment. If the user of
U
) for 5 cycles
T
the
Station
70% U
(30% dip in
T
U
) for 25 cycles
continued operation
T
during power mains
<5% U
(>95% dip
T
interruptions, it is
in U
) for 5 sec.
T
recommended that the
SurePower Charger
Station
an uninterruptable power
supply or a battery.
3 A/m
Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
commercial or hospital
environment.
-G
UIDANCE AND
SurePower Charger
unit requires
unit be powered by
9650-0535-01 Rev. C