Charger And Battery Keys; Battery Charging Requirements - ZOLL SurePower Operator's Manual

Charger station
Hide thumbs Also See for SurePower:
Table of Contents

Advertisement

Charger and Battery Keys

ZOLL's defibrillator batteries and the bays in the SurePower Charger Station are keyed to
prevent incompatible batteries from seating in the charging bays and damaging the SurePower
Charger Station. The following illustration shows the key on the SurePower Battery Pack.
Figure 1-3 Battery Key
Batteries that do not have the correct keys will not seat in the charging bays and cannot be
charged and tested by the SurePower Charger Station. For example, the AED PRO
defibrillator's disposable battery pack is keyed differently and will not seat in the SurePower
Charger Station.
Caution
Do not use the SurePower Charger Station to charge any battery packs which are not designed
for use with SurePower Charger Stations.

Battery Charging Requirements

Each battery type has its own charging requirements. When you place a SurePower Battery
Pack into the bay of a SurePower charger, the charger identifies the battery type and the battery
communicates its charging requirements to the charger.
9650-0535-01 Rev. C
SurePower Charger Station Operator's Guide
SurePower Charger Station Charging Bays
1-3