Displays During The Experiment; Remote Terminal Displays - Branson Digital Sonifier 250 User Manual

Hide thumbs Also See for Digital Sonifier 250:
Table of Contents

Advertisement

Digital Sonifier
Keypad Enabled
I f t h e k e y p a d i s e n a b l e d , t h e U p a n d D o w n a r r o w s , S t a r t / S t o p , a n d P a u s e k e y s a r e a c t i v e , a n d t h e d i s p l a y s h o w s
t h e i n f o r m a t i o n d e s c r i b e d i n t h e f o l l o w i n g t a b l e :
Table 6-4:
Data Displayed During the Experiment when the Keypad is Enabled
Display
RUN TIME
PROBE TEMP
PULSE ON and PULSE OFF
Power display bar graph
MAX. TEMP
PULSE TEMP
TIME
POWER
Keypad Disabled
I f t h e k e y p a d i s d i s a b l e d , t h e L C D d i s p l a y o n t h e f r o n t p a n e l s h o w s o n e o f t h e f o l l o w i n g i t e m s :
KEYPAD DISABLED message
EXPERIMENT RUNNING message
Power display bar graph
D u r i n g t h e e x p e r i m e n t , o n l y t h e S t a r t / S t o p a n d P a u s e k e y s a r e a c t i v e . W h e n t h e e x p e r i m e n t i s f i n i s h e d , t h e
E X P E R I M E N T R U N N I N G m e s s a g e c h a n g e s t o E X P E R I M E N T C O M P L E T E D .

Remote Terminal Displays

I f a r e m o t e c o m p u t e r o r t e r m i n a l i s c o n n e c t e d a n d a c t i v e , t h e s y s t e m d i s p l a y s t h e E X P E R I M E N T R U N N I N G
m e s s a g e o n t h e s c r e e n e v e r y f i v e s e c o n d s w h i l e t h e e x p e r i m e n t i s r u n n i n g . S e e p a g e 6 - 3 5 f o r m o r e i n f o r m a t i o n
a b o u t t h e R e m o t e T e r m i n a l o p t i o n a n d i t s c o m m a n d s .
100-214-239 Rev. A

Displays During the Experiment

The elapsed time of the experiment, increasing from 0:00:00
The temperature currently being read by the temperature probe, if one
is connected
If you entered Pulse On and Pulse Off times, shows PULSE ON when
ultrasonics are active and PULSE OFF when ultrasonics are inactive
The current percentage of maximum power being used, represented as
a bar graph. Displayed values are approximate, using 5% segments
The maximum temperature that will be reached during the experiment,
if you specified a Maximum Temperature parameter
In Pulse/Pause mode, the temperature at which the system will begin
calculating the Pulse/Pause ratio
The time limit for the experiment, if one was entered
The power, in watts, being consumed by the Horn at any moment. Dur-
ing a PULSE OFF period, the power goes to zero
Description
6-43

Advertisement

Chapters

Table of Contents
loading

This manual is also suitable for:

Digital sonifier 450

Table of Contents