Operation

Parameter Entry - Pulse Temperature

T h e P u l s e T e m p e r a t u r e i s t h e l o w e r t e m p e r a t u r e p a r a m e t e r o f t h e P u l s e / P a u s e m o d e t e m p e r a t u r e r a n g e .
E n t e r i n g a v a l u e f o r t h e P u l s e T e m p e r a t u r e e n a b l e s t h e P u l s e / P a u s e m o d e ; c l e a r i n g t h i s v a l u e d i s a b l e s t h i s
m o d e .
CAUTION
j
Setting the PULSE TEMP parameter to a numeric value will activate the Pulse/Pause Mode
process. This can cause the system to operate in an On and Off fashion, which can cause the
ultrasonics to be suspended for a period of time, and then reactivate to continue a process
cycle without warning.
Do not handle the Horn or Converter if the Pulse/Pause Mode is activated and running, and
ensure the process cycle is Stopped before touching the Horn, or injury could occur.
Take this action...
Use the Up and Down arrow keys, if
necessary, to display the line showing
the Pulse Temp parameter.
Press the Line Select key that corresponds
to Pulse Temp.
6-32
To get this result...
100-214-239 Rev. A

Advertisement