Change Parameters Screens; Cpap Mode; S Mode - Respironics BiPAP Synchrony Provider Manual

Hide thumbs Also See for BiPAP Synchrony:
Table of Contents

Advertisement

5-2

5.2 Change Parameters Screens

5.2.1
CPAP M
Parameter
CPAP
Ventilation Ramp
5.2.2
S M
ODE
S
PARAMETERS:
IPAP
19
19 cm H2O
EPAP
5
Rise Time
Screen
Number
Screen 1 of 2
IPAP
EPAP
Rise Time
Screen 2 of 2
Ramp
Bi-Flex
NOTE:
IPAP cannot be set lower than EPAP.
When Bi-Flex is enabled, Rise Time is set to 3 (0.3 seconds).
When Bi-Flex is enabled, then IPAP and EPAP are limited to 20 cm.
The Rise Time of 1 to 6 corresponds to tenths of a second (e.g., a setting of 4 equals 0.4 second rise time).
ODE
CPAP PARAMETERS:
CPAP
10
VentRamp OFF
4 to 20 cm H
Disabled, 0.5, 1.0, 2.0, 3.0 cm H
1/2
S PARAMETERS:
V entRamp OFF
5 cm H2O
B i-Flex
3
Parameter
4 to 30 cm H
4 to 25 cm H
1 to 6
Disabled, 0.5, 1.0, 2.0, 3.0 cm H
OFF, 1, 2, 3
R
10 cm H2O
EXIT
Range
O
2
O per breath
2
2/2
OFF
EXIT
Range
O
2
O
2
O per breath
2
Provider Manual
Increment
of Change
1 cm H
O
2
Discrete values as listed
Increment
of Change
1 cm H
O
2
1 cm H
O
2
1
Discrete values as listed
Discrete values as listed

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents