Medium Priority Alarms - Respironics BiPAP Synchrony Provider Manual

Hide thumbs Also See for BiPAP Synchrony:
Table of Contents

Advertisement

8-10
LED
Audible
Red Flash
Red Flash
Red Flash
Red Flash
1
Optional
2
The error codes can be viewed upon start-up of the Synchrony on the self-test screen. The codes can also be viewed by
displaying the System Codes screen as described in Section 7.2.2. Appendix B lists the error codes and their descriptions.
8.5.2
M
EDIUM
LED
Audible
Yellow
Flash
Yellow
Flash
Synchrony
Display
Action
Message
Low Pressure
Operates
Support
Low Minute
Operates
1
Ventilation
Ventilator
Shuts down
Inoperative
2
(Displays Codes)
Ventilator
Shuts down
Inoperative
Battery Failure
P
A
RIORITY
LARMS
Synchrony
Display
Message
Battery Voltage
Too High
Low Battery
R
Possible
Cause
Device
is malfunctioning.
Device sensed minute
ventilation below
alarm setpoint.
Internal device
failure.
Battery is
discharged.
Possible
Action
Cause
Operates
Wrong or
malfunctioning
battery or charger.
Operates
Battery nearly
discharged.
Provider Manual
Patient
Provider
Action
Action
Call your home
Replace the
care provider.
device.
Report alarm
Verify patient
to your home care
status.
provider.
Remove power
Replace the
from the device.
Synchrony.
Call your home
care provider.
Remove DC power
Replace the
source from the
battery.
Synchrony, replace
the battery, restore
power to the
Synchrony; or seek
reliable AC power
source.
Patient
Action
Remove DC power
Replace the
from the device,
battery.
use on AC, or call
home care provider.
If on battery, replace
Replace the
battery; if on AC,
battery.
exchange the battery
and charge.
Provider
Action