NOTE:
All differential cable pairs must be matched.
WARNING: V
must not be connected for NBSG16 evaluation board connection or damage may result
CTRL
Figure 2. NBSG16/16VS Board Setup - Time Domain (Differential Mode)
Setup (continued)
Setup Input Signals
Step 3:
3a: Set the signal generator amplitude to 500 mV.
NOTE:
The signal generator amplitude can vary from 75 mV to 900 mV to produce a 400 mV DUT
output.
3b: Set the signal generator offset to 660 mV (the center of a nominal RSECL output).
NOTE:
The V
vary as long as V
information.
3c: Set the generator output for a PRBS data signal, or for a square wave clock signal with
a 50% duty cycle.
Connect Output Signals
Step 4:
4a: Connect the outputs of the device (Q, Q) to the oscilloscope. The oscilloscope
sampling head must have internal 50 W termination to ground.
NOTE:
Where a single output is being used, the unconnected output for the pair must be terminated to
through a 50 W resistor for best operation. Unused pairs may be left unconnected. Since
V
TT
= 0 V, a standard 50 W SMA termination is recommended.
V
TT
NBSG16BAEVB, NBSG16VSBAEVB
V
= 2.0 V
CC
V
CC
D
D
V
V
BB
EE
V
BB
V
= -1.3 V (3.3 V op)
EE
V
= -0.5 V (2.5 V op)
EE
(Input High Voltage Common Mode Range) allows the signal generator offset to
IHCMR
is within the V
IH
IHCMR
http://onsemi.com
NBSG16VS
V
= 0 V
V
ONLY
TT
CTRL
GND
C2
C1
Q
Q
C1
GND
V
= 0 V
TT
range. Refer to the device data sheet for further
3
Tektronix TDS8000
Digital Oscilloscope
Channel 1 (80E01)
Channel 2 (80E01)
TRIGGER