Guidance and Manufacturer's Declaration – Electromagnetic
Immunity
The AVOXimeter 4000 is intended for use in the electromagnetic environment specified below.
The customer or the user of the AVOXimeter 4000 should assure that it is used in such an
environment.
Immunity Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical fast
transient / Burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions, and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
64
IEC 60601
Compliance
Test Level
Level
+ 6 kV contact
+ 6 kV contact
+ 8 kV air
+ 8 kV air
+ 2 kV for
+ 2 kV for
power supply
power supply
lines
lines
+1 kV for
+1 kV for
input/output
input/output
lines
lines
+1 kV
+1 kV
differential
differential
mode
mode
+2 kV
+2 kV common
common
mode
mode
<5 % U
T
<5 % U
(>95 % dip in
(>95 % dip in
U
for 0.5
T
U
T
cycle)
cycle)
40 % U
T
40 % U
(60 % dip in
(60 % dip in
U
) for 5
T
U
) for 5 cycles
T
cycles
70 % U
70 % U
T
(30 % dip in
(30 % dip in
U
) for 25
T
) for 25
U
T
cycles
cycles
<5 % U
<5 % U
T
(>95 % dip in
(>95 % dip in
U
) for 5 sec
T
U
) for 5 sec
T
3 A/m
3 A/m
Electromagnetic Environment -
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that of a
typical domestic, commercial or
hospital environment.
Mains power quality should be that of a
typical domestic, commercial or
hospital environment.
T
for 0.5
Mains power quality should be that of a
typical domestic, commercial or
T
hospital environment.
If the user of the AVOXimeter 4000
System requires continued operations
T
during power mains interruptions, it is
recommended that the AVOXimeter
4000 System be powered from an
uninterruptible power supply or battery.
T
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical domestic,
commercial or hospital environment.
Guidance