Plastic Tie
Plastic Tie
11.
Remove the battery, being careful not to snag any components when pulling the
cable out from underneath the main circuit board. If the battery well contains
tape, remove the tape.
12.
Remove the liner from the tape on the bottom of the new battery, and place the
new battery in the battery well with the tape side down, ensuring that the blue and
white cables are facing up, then route the new plastic ties (packaged with the
replacement battery) underneath the battery well.
New
Plastic Tie
New
Plastic Tie
AVOXimeter 4000 Operator's Manual
Blue and White
Battery Cable
New
Battery
55