Interferences
Fetal Hemoglobin
(tHb = 13.5 g/dL, HbF = 100%
Indocyanine Green Dye
(tHb = 16 g/dL, MetHb <10%,

Calibration

The AVOXimeter 4000 is factory-calibrated and employs highly stable state-of-the-art light
sources. Should recalibration be required please contact a Technical Support representative.
Proper calibration also requires entry of the correct cuvette pathlength by the user (see page 30)
and use of a customary value for Hüfner's number (see page 30).
CAUTION: If quality control results are not acceptable, erroneous results are
encountered, or error messages are displayed, the most likely cause may be
contamination of the optical detector by blood or debris, which cannot be resolved by
re-calibration. Consult the Troubleshooting section for additional information.