Specifying a Different Value for Hüfner's Number
Hüfner's number (the volume of oxygen that can be carried by one gram of hemoglobin) is used
to calculate the oxygen content of a sample (see page 6). Although Hüfner's number is generally
assumed to be 1.39 mL/gm, facilities may use a different value and may wish to set the
AVOXimeter 4000 to match other instruments in the facility.
1.
2.
3.
4.
5.
6.
30
Display the "Calibration" menu (a submenu of
Press
followed by
value used for Hüfner's number:
Press
followed by
Note:
A default value of 1.39 is entered at the factory. Values ranging from
1.30 to 1.39 can be entered.
Press the number key (e.g.,
Check that the new Hüfner's number is correctly displayed, then press
followed by
. The "Calibration" menu is again displayed.
Press
to return to the previous menu, if desired.
, see page 12).
to display the screen for accepting or changing the
to display the screen for entering a different value:
) corresponding to the new value: