Advertisement

AVOXimeter 4000 Operator's Manual
4.
5.
6.
7.
8.
9.
10.
11.
40
The screen for specifying the sample type is displayed:
Press
followed by
Press
followed by
Press the number key corresponding to the filter being run, followed by
Press
followed by
displayed, and test results are displayed within ten seconds.
Record the THb, %O
Hb, %COHb, and %MetHb results, or print the results of the
2
test (see page 36).
Press any key to clear the results, then repeat steps 2 through 8 using the orange
filter.
Verify that the results for each filter are within the expected ranges listed below:
Optical Filter
tHb (g/dL)
Yellow
7.8 to 8.2
Orange
16.7 to 17.3
Note:
The Expected Range is also shown on a sticker on the side of each filter.
If desired, plot the optical filter results on daily logs such as the ones shown in the
Quality Control Logs section on page 57.
. The screen for specifying QC type is displayed:
. The screen for specifying filter type is displayed:
to confirm your selection. A confirmation screen is
Expected Range
%O
Hb
2
93.7 to 96.3
37.8 to 40.2
%COHb
%MetHb
0.6 to 3.4
-0.4 to 2.4
20.0 to 23.0
0.2 to 1.8
.