TOOUBLESHOOTING GUIDE
Issue
No LCD display
No rate or total
displayed
Flow rate display
interprets reading
constantly
Flow rate
indicator
bounces
PAOT NUMBEOING INFOOMATION
B 28 S X X - X X
Program Level
S -
Power Option
B - Battery
L - Loop
March 2017
•
Check battery voltage� Should be 1�5V DC� Replace if low or bad�
•
Check connection from meter pickup to display input terminals�
•
Check turbine meter rotor for debris� Rotor should spin freely�
•
Check programming of flow monitor�
•
This is usually an indication of external noise� Keep all AC wires
separate from DC wires�
•
Check for large motors close to the meter pickup�
•
Check for radio antenna in close proximity�
•
Try disconnecting the pickup from the monitor pig tail� This
should stop the noise�
•
This usually indicates a weak signal� Replace pickup and/or check
all connections�
•
Check K-factor�
DSY-PM-00276-EN-04