Operator Maintenance; Replacing The Power Line Fuse; Replacing The Fuse - HP E4418B User Manual

Power meter
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

Operator Maintenance

This section describes how to replace the power line fuse and clean the
power meter. If you need additional information about replacing parts or
repairing the power meter, refer to the HP E4418B/E4419B Service
Guide.
To clean the power meter, disconnect it's supply power and wipe with a
damp cloth only.
The power line fuse is located within the power meter's fuse holder
assembly on the rear panel. For all voltages the power meter uses a 250 V,
F3.15AH, 20mm fast blow fuse with high breaking capacity (HP part
number 2110-0957).
Note

Replacing the Power Line Fuse

HP E4418B User's Guide
The power meter also has an internal fuse. If you suspect that this
fuse needs replaced it must be done by trained service personnel.
Please refer to "Returning Your Power Meter for Service", on
page 2-80.
1. Remove the power cord from the power meter.
2. Slide the fuse holder assembly from the rear panel as shown in
Figure 2-23.
3. Install the correct fuse in the "In line" position as shown in
Figure 2-23. (A spare fuse can be stored in the fuse holder
assembly.)
4. Replace the fuse holder assembly in the rear panel.
Figure 2-23: Replacing the Fuse
Power Meter Operation
Operator Maintenance
In line fuse
Spare fuse
2-73

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents