Front Panel Menu Reference "System/Inputs" Menu
Echo
Press this softkey to toggle Echo "On" or "Off". When Echo
is enabled, received characters are transmitted back to the
sender.
Tables
Press this softkey to access a menu which allows you to
select between sensor calibration tables and frequency
dependent offset tables.
Sensor Cal Tables
Press this softkey to access a menu and screen which allows
you to select and edit sensor calibration tables. Use the
and
calibration tables. Sensor calibration tables are only required
for the HP 8480 series power sensors.
Edit Table
Press this softkey to access a menu and screen which allows
you to modify, add and remove frequency, calibration factors
and offsets from the currently selected table and modify the
table name. Use the
move between the name, frequency, calibration factors and
offsets. Press
Change
Press this softkey to modify the highlighted parameter,
which is either a frequency, calibration factor, offset or
table name. A pop up window displays the parameter. Use
the
parameter value. To confirm your choice press
Insert
Press this softkey to insert a new table entry. You are
prompted to enter the frequency and calibration factor or
offset. The entry is inserted in ascending frequency order.
Delete
Press this softkey to delete a table entry. If you delete the
frequency the corresponding calibration factor or offset is
also removed and vice versa.
Done
Press this softkey to return to the measurement screen.
The softkeys from the first level of the "System/Inputs"
menu are displayed.