Power Meter Operation
Making Measurements with the HP E-Series Power Sensors
Example
To make a measurement using an HP E-series power sensor. The
frequency of the signal to be measured is 100 MHz.
•
Disconnect the power sensor from any source.
Zero
•
Press
.
Cal
•
Press
.
Zero
•
Connect the power sensor to the POWER REF output.
•
Press
.
Cal
Frequency
•
Press
Cal Fac
hardkeys to enter 100. Press
•
Connect the power sensor to the signal to be measured.
•
The measurement result is now displayed.
2-16
,
. Use the
Freq
.
MHz
,
,
and
HP E4418B User's Guide