Power Meter Operation
Presetting the Power Meter
Presetting the Power Meter
This section details the power meter's preset conditions.
The HP-IB address and command set, the data stored in the sensor
calibration tables and the zeroing and calibration data are not affected by
a preset. The calibration table selected is not affected.
Preset Conditions
The number of windows displayed is two.
dBm/W
Measurement units used is set to dBm.
Frequency
Cal Fac
•
is set to 50 MHz.
Freq
•
is set to 100%.
Cal Fac
Meas
Setup
•
Display Format
analog in the lower window.
•
Anlg Mtr Scaling
(
).
Min
•
Resolution 1 2 3 4
•
Limits Off/On
•
is set to 90.000 dbm (
Limits
•
TTL Output Off/On
•
Limits OVER UNDER EITHER
•
Fail O/P High Low
Rel
Offset
•
Rel Off/On
•
Offset Off/On
•
is set to 0.000 dB.
Offset
System
Inputs
2-64
is set to digital in the upper window and
is set to 20.000 dBm (
is set to "3".
is set to "Off "
) and -90.000 dBm (
Max
is set to "Off ".
is set to "OVER".
is set to "Low"
is set to "Off".
is set to "Off ".
) and -70.000 dBm
Max
).
Min
HP E4418B User's Guide