Calibration by Certificate
In some circumstances it may not be possible to apply inputs in which case the calibration can be entered
manually from the calibration table or certificate for a load cell without ever having to connect the load cell.
Items you can change:
Number of Calibration Points
Input Points 1 – 9
(mV/V shown in this
screenshot)
Engineering Units 1 - 9
Calibrate
Mantracourt Electronics Limited
Enter the number of points you wish to calibrate over. In its simplest form you
could select two for a linear calibration.
For more complex calibrations which include linearization select three to nine
points.
Enter the input point for which you will specify a required engineering output
value
Enter the required engineering unit output for the specified input value
Click this button to calculate and update the module calibration
47
T24 Telemetry User Manual