Input / Output Configuration

This module does not provide calibration, as such, because it is factory calibrated. However, on this page you can
select the output type and parameters unique to your input sensor.
Items you can change:
Output Type
Pulses per Revolution
Edge Type
Debounce Filter
Ignore First Pulse
Mantracourt Electronics Limited
Simply select the required output type from the drop down list.
Frequency (Hz) – Average frequency of pulses measured over the sample time
Time (s) – Average time in seconds between pulses measured over the sample
time
RPM – Average Revolutions Per Minute measured over the sample time.
Counter – Counts incoming pulses. Count is reset by digital input to GND or
external reset using data provider packet .Should not be used in low power
mode
Specify the number of pulses per revolution. This parameter only affects the
RPM output value.
Define which edge of an input pulse should be counted as the input trigger
Enter a time in milliseconds, any pulse that is received within this time of a
previous pulse will be ignored, this is useful when dealing with noisy inputs such
as relays which may inadvertently produce more than one pulse per event.
If set to yes the T24-PA will ignore first pulse received during the sample time,
before continuing to average the time between the subsequent pulses. This is
useful in situation where a sensor may be powered by the transmitter module
and may produce an erroneous pulse on start up.