If required, apply pressure of the day correction (POD), by selecting POD ADJUST
as follows:
SETUP
QFE pressure reading as POD correction as follows:
Apply
Note: The display can show either: POD = Off or POD = current value (QFE pressure
reading). The POD RESET key (F2) sets the current POD correction to Off.
POD = Off
POD Correction
WARNING
VENT SYSTEM
BEFORE SETTING POD
POD RESET
CHANGE POD
VALUE
Method (accessible switch contacts)
(1)
Connect the instrument to the static system and connect the altitude switch as
shown. Ensure that the pitot system is vented to atmosphere before starting the test:
Contacts must be potential free.
Note:
Figure C3 - Altitude Switch Test Connections
(2)
Select negative pressure and close the release valve.
K0237 Issue No. 2
SETUP
BACKLIGHT
STORE MODE
CONTRAST
SETTINGS
DRY CELL LEVEL
POD = Off
POD Correction
mbar
WARNING
VENT SYSTEM
BEFORE SETTING POD
CHANGE POD
VALUE
Applied Static System Pressure
Vent Pitot System (Atmospheric Pressure)
Select
Negative (-)
Pressure
SELECT OPTION
SETTINGS
UNITS
CALIBRATION
LANGUAGE
RS232
POWERDOWN
CALIBRATION
POD ADJUST
ENTER VALUE
mbar
POD RESET
SWITCH TEST
CONTACT STATE
POD = 1008.63
ALTITUDE
RUN
F1
4
SELECT OPTION
POD = 1002.85
POD Correction
WARNING
VENT SYSTEM
BEFORE SETTING POD
CHANGE POD
VALUE
Altitude Switch
S
ft
F2
mbar
POD RESET