(3) Functions of test mode
1. Minute temperature display
When displaying the current temperature, the
value after the decimal point is also displayed.
2. Expansion of temperature setting range
Temperature can be set from 20"C to 150"C.
Error is prohibited by the sterilization limiter.
1
Displays 112.3"C.
Set value can be changed even after heating process.
3. Remaining time display in exhaust process
Exhaust remaining time is displayed on the Digital Display II. (in min.)
4. Pressure/Saturated steam pressure display
1-
, L
I
By pressing the "TIMER" button, the current pressure is the Digital Display I. (in kPa)
5. Forced exhaust
12°31
I
By pressing the
"A"
and 'STOP" buttons at the same time during stand-by, the exhaust valve opens.
The forced exhaust can be cancelled by pressing the "STOP" button.
6. TImer 60-time speed operation
While the "TIMER" and the "SET/CONFIRM" button at the same time, the reservation timer, process
timer and sterilization timer operate in 60-time speed. (It is preferable to perfonn the exhaust fast.)
7. sterilization process starUcompletion signal output
Buzzer sounds at the start and completion of the sterilization process.
(4)Functions of demonstration mode
1. Power is not supplied to the heater.
2. Process timer and sterilization timer operate in 60-time speed.
3. When the power is supplied to the heater, the temperature on the Digital Display rises automatically.
4. When the power is not supplied to the heater, the temperature on the Digital Display lowers automatically.
(For safety sake, however, the temperature does not lower than the actual temperature of TH1.)
5. Pressure Abnonnal is not detected.