Download Print this page

Sanyo MLS-3751L Service Manual page 30

High-pressure steam sterllizer labo autoclave

Advertisement

1
Press the SET LCONF I RII button
to
set the sterilization temp. at
121''c.
Reaching 121
Perform the 121"C correction in
the same manner as the
correction at 1 05"C.
Press the SET LCONF I RII button
to
set the sterilization temp. at
135"C.
Reaching 135
Perform the 1350C correction in
the same manner as the
correction at 105"C.
Set
the sterilization temp.
"C
and sterilization process
Adjust the temp. to 121
"C
after
going
to
the 121
etc
sterilization.
Set the sterilization temp.
"C and sterilization process
Adjust the temp. to 135"C after
going
to
the 135"C sterilization.
- 29-
Example
of
the lateral thermistor correction
display (in 121"C correction)