Sanyo MLS-3751L Service Manual page 27

High-pressure steam sterllizer labo autoclave

Advertisement

@Exhaust setting
-Heater OFF temperature can be changed by pressing the" ."
or "T" button.
(Heater ON temperature is fixed at the boiling point temp.+1
"C.)
-When the" ."or "T" button is pressed while pressing the
"SET/CONFIRM" button, the exhaust time can be changed.
-When the "START" button is pressed while pressing the
"SET/CONFIRM" button, the content of the change is written
in the non-volatile memory.
<!lAddress
-Address can be changed by pressing the" ."or "T" button.
This address is used for connection of the MTR system.
-When the "START" button is pressed while pressing the
"SET/CONFIRM" button, the content of the change is written
in the non-volatile memory.
@Option 1
-Change bit is selected by pressing the" • "or "T" button.
-Preset value can be changed by pressing the "TIMER" button
-When the "START" button is pressed while pressing the
"SET/CONFIRM" button, the content of the change is written
in the non-volatile memory.
@Optlon 2
-Change bit is selected by pressing the" • "or "T" button.
-Preset value can be changed by pressing the "TIMER" button
-When the "START" button is pressed while pressing the
"SET/CONFIRM" button, the content of the change is written
in the non-volatile memory.
@lCH display ferquency
CH display frequency is set. (CH is displayed when the total
processing frequency exceeds the CH display frequency.)
4 kinds of frequency of 1000 .. 2000 .. 3000 (Factory setting)
.. 9999.4 can be selected.
-When the "T" button is pressed, the frequency decreases and
when the"." button is pressed, the frequency increases.
-When the "START" button is pressed while pressing the
"SET/CONFIRM" button, the content of the change is written
in the non-volatile memory.
«Note» Do not set 9999.
- 26-
RP[j1
B 1
Heater OFF temp.
SR
o
OP ,
lFbO
i
I
Change bit
I
I
Preset value
I
OP2
FFbO
i
11..)-_---,
I
Preset value
I
I
Change bit
I
[H
3000