(11) Abnonnal teQJerature of the bottom thennlstor (TH3) (E5 occurs.)
1 Check
the
heBlin water amount
in
the chamb...
If
the heati
water In the
chamber
is d .... Bled, re enish it
2
Check.
the
nem of "steam leaks from
I e and oint."
Re
Ir the
steam
leakl
rl
. ,
.... +-
SolenoId
Between White 2 and 3
connection.
Housing set DS
Detection
switch
Conductivity checking
posilions
• DetecUon switch 1
Between White 2
.oo
3
*
Wire Nos. are
indicated on
the
connectors.
I
(1) Check connector insertion
of
the housing
set
X+ U(3751)
or D
+
T(3781). If not completely Inserted,
correct
n.
(2) Check
the
resistance
value between the terminals of
the
solenoid.
Range of resistance value: 160 - 1852
In
cue
of
short-an::uit
or breakage, replace
the
solenoid.
(3) Check. If the lock pin Is caught. Apply grease.
incomplete
short-circuited
with
housing
set
DS.
lever being retumed, replace
the
- 41 -