@Back side exhaust tank
When the water temperature in the exhaust lank rises
to
make it impossible to condense steam to water,
the steam is exhausted into the back side exhaust lank.
It is the back side exhaust lank that recovers the steam.
(Cautions)
When connecting the exhaust hose to the back side exhaust lank, be sure to insert it so that its end
reaches near the bottom of the back side exhaust lank.
The water in the back side exhaust lank returns from the back side exhaust lank to the exhaust
lank. In case
of
incomplete connection
of
hose, water remains in back side exhaust lank and the
water may splash in all directions in the exhaust process.
@Thermostat
This is the final safety device to prevent abnormal temperature rise due
to
heating of empty chamber by
heater. The power supply is shut down when this device works. The thermoslat is of manual reset type. The
actuating temperature is 145°C±5°C.
The thermoslat is mounted to the bottom of the chamber. To reset it, remove the unit side panel and the
reset button must be pressed directly.
(Caution)
Since the thermoslat can not be rest at a high temperature, rest it when the chamber is cooled well.
@Chamber lid packing
Timing of the chamber lid packing replacement is 1 year as slandard. If deterioration
of
the edge, hardening,
remarkable deformation or lowering of elasticity is found, it must be replaced.
When replacing, tum OFF the power supply and then remove the chamber lid packing from the chamber
lid.
Make sure that no foreign matters such as dirt, etc. are sticking
to
the new lid packing and the lid before
mounting the packing
to
the chamber lid.