Fluke 52120A User Manual page 79

Transconductance amplifier
Hide thumbs Also See for 52120A:
Table of Contents

Advertisement

5720A
Output
-2 Vdc
0.56 V, 10 Hz
1.4 V, 10 Hz
0.8 V, 57 Hz
2 V, 57 Hz
0.8 V, 300 Hz
2 V, 300 Hz
0.8 V, 1 kHz
2 V, 1 kHz
0.8 V, 3 kHz
2 V, 3 kHz
0.8 V, 6 kHz
2 V, 6 kHz
0.8 V, 10 kHz
2 V, 10 kHz
20 Amp Range Verification
1. Push S on the Product and the 5720A.
2. Connect the 20 A shunt between the output of the Product and the DMM front input.
3. Set the DMM to front input.
4. Push C on the Product.
5. Set the 5720A output to +1 V dc.
6. Push M on the 5720A and the Product.
7. Let the measurement become stable for a minimum of one minute and then record the
measured error in Table 5-3.
8. Set the DMM 'm' store to the 20 A shunt correction factor.
9. Set the DMM '-c' and '÷z' stores to 0.4.
10. Turn on the '*m', '-c' and '÷z', and % math functions on the DMM.
11. Set the DMM to front input.
12. Set the DMM '-c' and '÷z' stores to 0.4.
13. Set the 5720A output to -1 V dc.
14. Set the DMM '-c' and '÷z' stores to 0.8.
15. Turn on the '*m', '-c' and '÷z', and % math functions on the DMM.
16. Set the 5720A output to the values in Table 5-4 for the +2 V dc and -2 V dc
Table 5-3. 2 Amp Range Verification (cont.)
8508A
8508A
ACV
A40B
Input
filter
-
Front
2 A
10
Rear
2 A
10
Front
2 A
40
Rear
2 A
40
Front
2 A
100
Rear
2 A
100
Front
2 A
100
Rear
2 A
100
Front
2 A
100
Rear
2 A
100
Front
2 A
100
Rear
2 A
100
Front
2 A
100
Rear
2 A
100
Front
2 A
Product
Measured
Specification
Error (%)
(99 %)
±0.0150 %
-
Reference level
±0.1150 %
-
Reference level
±0.0850 %
-
Reference level
±0.1000 %
-
Reference level
±0.1700 %
-
Reference level
±0.9000 %
-
Reference level
±0.2600 %
-
Reference level
±6.0000 %
Calibration
Introduction
Measurement
uncertainty
TUR
(99 %)
±0.0030 %
5.0:1
±0.0051 %
23:1
±0.0037 %
23:1
±0.0037 %
>25:1
±0.0037 %
>25:1
±0.0040 %
>25:1
±0.0040 %
>25:1
±0.0040 %
>25:1
5
5-9