Figure 2-2. Product Connections to a Watthour Meter
The capacitance between the measurement circuits and the grounded meter case is too
small to pass a common-mode signal of sufficient amplitude to change the measurement.
Devices with their own power supplies which have an isolation input transducer (such as
current transformers) are relatively immune to common mode current errors. The path to
ground is interrupted by the transducer.
Line Side
N
Vin
Shunt
Resistor
N
Load Side
Front-Panel Operation
Notes on Product Use