Transconductance amplifier
Hide thumbs Also See for 52120A:
Table of Contents

Advertisement

CALibration:SECure:PASSword <spd>
Description Sends a password to the Product to put the Product into adjustment mode.
The calibration switch on the rear panel of the Product must be set to
Enable before you send the password.
Parameters <spd>
Example
CAL:SEC:PASS "ADJUST"
CALibration:SECure:EXIT
Description Exits the calibration or adjustment mode.
Example
CAL:SEC:EXIT
All changes will be lost the next time the Product is power cycled, unless
CAL:SAVE? is done before the exit command.
CHAin:FITTed?
Description Gets the number of units in the chain.
Example
CHA:FITT?
CHAin:IDN? <nrf>
Desription
Gets the identification string of an instrument in the chain.
Parameter
<nrf>
Example
CHA:IDN? 2
The command
Returns
<manufacturer>, <model>, <serial number>, <firmware>
INPut:TYPE(?) <cpd>
Description Sets or gets the input type selection.
Parameter
<cpd>
Example
INP:TYPE CURR
INP:TYPE?
OUTPut[:STATe](?) <bool>
Description Sets or gets the state of the output signal.
Parameters <bool> =
Example
OUTP ON
Note
=
(the password)
Exits calibration adjustments.
Note
Get back a number from 1 to n.
=
1 to 10
Index of the instrument in the chain.
Gets back the identification string of the second
instrument in the chain.
Note
is equivalent of
:CHA:IDN? 1
=
CURRent
Sets the input type to current.
VOLTage
Sets the input type to voltage.
Sets the input type to current.
Gets back
type is set in the Product.
ON or 1
Enables the output signal (OPR).
OFF or 0
Disables the output signal (STBY).
Turn the Product output on.
Sends the string
ADJUST
allow calibration adjustments. Note the word
ADJUST must be in quotes for this
command.
.
*IDN?
or
depending on which
CURR
VOLT
Remote Operations
Remote Commands
to the Product to
3
3-9