How To Connect The Product To Mains Power - Fluke 52120A User Manual

Transconductance amplifier
Hide thumbs Also See for 52120A:
Table of Contents

Advertisement

52120A
Users Manual
To increase the life of the Product:

How to Connect the Product to Mains Power

To prevent possible electrical shock, fire, or personal injury:
When exposed to low temperatures for an extended time, such
as air travel or storage, condensation may form inside the
Product. To prevent damage to the Product, let it acclimate to
its environment out of its shipping container a minimum of one
hour before you connect it to mains power.
The Product automatically senses mains voltage between 100 and 240 volts. No line
voltage or fuse selection is necessary. See the Maintenance section to learn how to
replace the mains fuse.
Because the Product can pull more current than a standard 10 A IEC connector, the
Product has a 16 A power connector on the rear panel. A mains power cord with 16 A
capacity is also supplied with the Product. Table 1-3 is a list of power cord types
available from Fluke.
1-8
Keep the area around the air filter a minimum of 4 inches from nearby walls or
rack enclosures.
Keep the inlet and exhaust perforations on the sides of the Product clear of
blockages.
Keep the air that goes in to the Product between 5 °C and 35 °C.
Make sure exhaust from a different instrument is not pointed into the fan inlet.
Clean the air filter at a maximum of 30 day intervals. More frequently if the
Product is operated in a dusty environment. See the Maintenance section of this
manual for instructions on how to clean the air filter.
Use only the mains power cord and connector approved for
the voltage and plug configuration in your country and rated
for the Product.
Replace the mains power cord if the insulation is damaged
or if the insulation shows signs of wear.
Make sure the ground conductor in the mains power cord is
connected to a protective earth ground. Disruption of the
protective earth could put voltage on the chassis that could
cause death.
Do not disconnect or open the protective ground conductor
inside or outside the Product. An open ground conductor
can make the Product dangerous.
XW Warning
W Caution