Mounting - Honeywell Sensepoint Technical Manual

Gas detector
Hide thumbs Also See for Sensepoint:
Table of Contents

Advertisement

Note
Diagram not to
scale

Mounting

Caution
The enclosure must be fixed in position using the external mounting
lugs and not mounted using any type of cable entry device.
Sensepoint should be fitted to a flat firm surface suitable for the unit's size and weight, e.g. a wall,
as follows:
1
Drill two mounting holes in the surface.
The surface should be flat and firm. For sizes, position, etc. see the previous dimension
drawing.
2
Mount the detector using its two mounting lugs.
Use M6 screws/bolts. Mount the unit with the gas sensor pointing downwards to ensure that
dust/water will not collect on the sensor, and to stop gas entering the detector.
2106M0525 Iss 2 Nov 05
76.5 mm (3.0")
Enclosure
Sensor
Sensor
Cap
34.0 mm (1.3")
Flammable
195.0 mm (7.7")
Toxic/Oxygen
205.0 mm (8.1")
Installation
12