Download Print this page

Electrical Specifications - AEMC MN375 User Manual

Ac current probe

Advertisement

2.56"
(65 mm)
.83"
(21 mm)

ELECTRICAL SPECIFICATIONS

Nominal Range: 10A
Measurement Range: 0.1 to 10A
Transformation Ratio: Voltage output
Output Signal: 100mV/A (1 V at 10A AC)
Accuracy and Phase Shift*:
Accuracy:
0.1 to 1A: 1% Reading ± 2mV
1 to 10A:
1% Reading ± 1.5mV
Shop for AEMC products online at:
2.00"
(51 mm)
ShopAEMC
www.
.73"
(18.5 mm)
.1.18"
(30 mm)
Phase Shift:
≤ 1° @ 60 Hz
1 to 5A:
≤ 1.5° @ 60 Hz
5 to 10A:
(*Reference conditions: 20°C to 26°C, 20 to 75% RH,
external magnetic field <40 A/m, 48 to 65 Hz sine wave,
distortion factor less than 1%, no DC component, no
external current carrying conductor, test sample centered.)
Load impedance > 1MΩ.
Frequency Range: 40 to 3kHz
Limit Operating Conditions:
20A permanently to 1kHz