Connecting The Pressure Gauge - Millipore Pellicon 2 User Manual

Table of Contents

Advertisement

Pellicon-2 and 3 Mini Holder User Guide
Figure 6:

Connecting the Pressure Gauge

Analog and digital pressure gauges are available from Millipore.
1.
2.
Figure 7:
Clamping the Pressure Gauge Adapter Fittings to the Mini Holder Fitting
Seat the ¾-inch sanitary gasket in the groove on the pressure gauge adapter fittings.
Hold the gasket in place and press the pressure gauge diaphragm against the ¾-inch
side branch of the pressure gauge adapter fittings, as shown in Figure 7.
Attaching the Pressure Gauge to the Pressure Gauge Adapter Fittings
9

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pellicon 2 mini holder

Table of Contents