Pellicon-2 and 3 Mini Holder User Guide
3.
Note The fi lter will extend beyond the edges of the manifold and end plate if the fi lter
4.
5.

Attaching the End Plate

1.
Figure 9:
2.
3.
4.
Align the filter cut-outs with the tie rods, and insert the filter so it is parallel with the
manifold as, shown in Figure 8.
does not properly align to the feed, retentate, and permeate holes.
Holding the manifold with one hand, slide the filter gasket (Pellicon 2 Cassettes
only) tabs over the tie rods and press the filter gasket (Pellicon 2 Cassettes only)
against the filter, as shown in Figure 8.
Repeat steps 3–5 for additional cassettes.
Holding the manifold gaskets (Pellicon 2 Cassettes only) and filter(s) in place with
one hand, slide the end plate holes over the tie rods, as shown in Figure 8, and press
the end plate against the filter.
Attaching the end plate
Place an equal number of spacers on each tie rod depending on the number of
filters in use.
Place one washer and one nut on each tie rod, as shown in Figure 8, and hand
tighten.
Tighten the holder with the torque wrench and socket. The torque wrench setting
for 180-200 inch-pounds (20.3-22.6 Newton-meters). Check the torque wrench
(not supplied with the holder) setting for 180-200 inch-pounds (20.3-22.6 Newton-
meters) prior to use. Refer to the next section to set the torque wrench.