Connect the Permeate Fittings

Two gaskets and two clamps are supplied for attaching fittings to the two PERMEATE
fittings located on the sides of the manifold plate. The PERMEATE fittings accommodate
standard ½-inch or ¾-inch sanitary fittings and adapters.

Inserting the Pellicon Cassettes

Use one to three cassettes depending on the desired permeate flow rate. Permeate flow
rate increases as filter area increases.
1.
2.
Note Pellicon 2 Cassettes ONLY
Figure 8:
Secure the pressure gauge to the Pressure Gauge Adapter Fittings with the sanitary clamp.
Open the sanitary clamp by moving the swing bolt out of the clamp slot.
Place the sanitary clamp over the Pressure Gauge Adapter Fittings-PRESSURE GAUGE
connection and close the sanitary clamp by moving the swing bolt into the clamp slot.
Hand tighten the sanitary clamp closure.
Repeat steps 1-3 to connect a Pressure Gauge Adapter Fittings to the
RETENTATE fitting.
Inspect the tie rods and nuts for signs of burrs or stripped threads.
turn freely on the tie rods for proper tightening of the holder.
Holding the manifold with one hand, slide the filter gasket (Pellicon 2 Cassettes only) tabs
over the tie rods and press the filter gasket against the manifold, as shown in Figure 8.
The large feed and retentate holes on the fi lter gasket must align to the large feed
and retentate holes on the manifold.Millipore supplies two fi lter gaskets with each
fi lter. One fi lter gasket must be installed between each pair of fi lters, and one gasket
must be installed between each end of the fi lter and the adjacent stainless steel plate.
Inserting the Filter Gaskets and Filter