Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The MicroCO is intended for use in the electromagnetic environment specified
below. The customer or the user of the MicroCO should assure that it is used in
such an environment.
Conducted
3 Vrms
RF
150 kHz to
IEC61000-4-
80 MHz
6
3 V/m
80 MHz to
Radiated RF
2.5 GHz
IEC61000-4-
Portable and mobile RF
communications equipment
should be used no closer to any
part of the MicroCO, including
any cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
3 Vrms
d = 1.2√P
3 V/m
d = 1.2√P
800 MHz
d = 2.3√P
2.5 GHz
Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and
is the recommended separation
distance in meters (m).
Fields strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey,
should be less than the
compliance level in each
frequency range.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
68
d
80 MHz to
800 MHz to
b