Disinfection of contaminated parts is only effective after having them
carefully pre-cleaned. CareFusion recommend the tested solution of
PeraSafe sterilizing powder (36-SSC5000A) for pre-cleaning and
disinfection. If a different solution is used, please follow the given
manufacturer's instructions.
Switch off the device and always unplug your MicroCO
meter from the computer before cleaning.
When using the disposable cardboard mouthpiece with one-way valve
under the prerequisite that the patient was instructed only to exhale,
the mouthpiece adapter and the CO meter surface have to be cleaned.
The mouthpiece adapter may be cleaned and sterilised by immersion
in Perasafe (36-SSC5000A) or other cold sterilising solutions. Exposed
surfaces of the meter may be wiped with a clean, damp cloth or a
cloth that has been immersed in a sterilising solution. It is
recommended that this procedure is performed after each use and all
used cardboard mouthpieces discarded.
aqueous solutions and
or permanent damage may result
Do not attempt to wash or immerse the MicroCO meter in
water or cleaning fluid, as there are electronic components inside that
will be permanently damaged.
immediately after the use.
If there are changes on the material surfaces of either the unit or
mouthpiece adaptor (cracks, brittleness) the respective parts must be
disposed off.
The sensor surface must not be wiped with any
Cardboard mouthpieces must be disposed off
be exposed to solvents e.g. alcohol