The MicroCO is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the
MicroCO can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment
(transmitters) and the MicroCO as recommended below, according to the
maximum output power of the communications equipment.
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures objects and
people
d
0.12
0.38
1.2
3.8
12
d
0.12
0.38
1.2
3.8
12
35
d
0.23
0.73
2.3
7.3
23.3