CareFusion MicroCO Operating Manual page 19

Table of Contents

Advertisement

adapter into the MicroCO meter and then insert a disposable
mouthpiece into the adapter. Turn the unit on by selecting the upper
or middle position on the central slide switch and the display will show
the software version number.
This will appear momentarily whilst correction for ambient levels is
executed. During this time the unit must not be exposed to raised
concentrations of CO. The buzzer will then sound and the display will
change to:
Upon hearing the buzzer instruct the subject to inspire fully and hold
their breath for 20 seconds. The display will count down from 20 to 0
as an aid to timing the breath holding period. The green indicator will
then illuminate and the display will change to:
BLOW
At this time the subject should seal their lips around the mouthpiece
and exhale slowly and fully. The 20 second breath holding period is
recommended to allow time for equilibration of alveolar gas.
If, however, the subject is unable to hold their breath for this period,
the unit may still be used before 20 seconds have elapsed.
17

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents