LaMotte smart spectro 2 Operator's Manual page 171

Spectrophotometer
Table of Contents

Advertisement

PROCEDURE
Use universal sample holder.
1.
Press ON button to turn spectrophotometer on.
2.
Scroll to and select PROGRAMMED TESTS from menu.
3.
Scroll to and select ALL TESTS (or another sequence containing
35 Cyanide) from TESTING MENU.
4.
Scroll to and select 35 Cyanide from menu.
5.
Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
6.
Dip the end of plastic rod (0519) into water sample and touch it to a small piece (1/4
inch) of pH test paper (2955) to wet paper. Read pH immediately from color chart.
a) If pH is below 10, raise the pH by adding *Sodium Hydroxide, 1N (4004) one drop
at a time with stirring. Check pH after each drop with a new piece of pH test paper.
Continue adjustment until pH is between 10.5 and 11.0.
b) If pH is above 11.5, lower pH by adding *Hydrochloric Acid (6130) one drop at a time
with stirring. Check pH after each drop with a new piece of pH test paper. Continue
adjustment until pH is between 10.5 and 11.0.
7.
Insert tube into chamber, close lid and select SCAN BLANK.
8.
Remove tube from Spectro. Use the 1.0 mL pipet (0354) to add 1.0 mL of Cyanide
Buff er (2850PS) to tube. Cap and mix.
9.
Use one 0.1 g spoon (0699) to add one level measure of *Cyanide Cl Reagent
(2794DS). Cap and invert 10 times to mix. Wait 30 seconds.
10. During the 30 second waiting period, carefully fi ll a second 0.1 g spoon (0699) with
one level measure of Cyanide Indicator Reagent (2793DS).
11. At the end of the 30 second waiting period, immediately add the level measure of
*Cyanide Indicator Reagent (2793DS). Cap and shake vigorously for 20 seconds. Wait
20 minutes for maximum color development.
12. At the end of the twenty minute waiting period, mix, insert tube into chamber, close lid
and select SCAN SAMPLE. Record result.
13. Press and hold OFF button to turn spectrophotometer off or press EXIT button to exit
to a previous menu or make another menu selection.
SMART Spectro 2 Test Procedures 05.02.16