EN-20
To switch between Over/Under Load, Target Weight
Tolerance, Target Weight Percentage press the button
corresponding to the
by using the alphanumerical keypad and press the button
corresponding to the
return to the previous screen.
Note: The three set limits methods share the same data.
4.3.2

Check Counting

Press the configuration button
Under/Accept/Over status is shown in the progress bar area while the actual number of pieces is shown on the
main Display Line.
Note: Refer section 9.5, or press the
Set the Average Piece Weight (APW)
Note: It is recommended that the APW is larger than 1d. If APW
is between 0.05d and 1d, a warning screen will be displayed and
the information line will show 'Low APW'. If APW is less than
0.05d an error screen will appear and the APW value cannot be
stored.
There are three ways to set the APW, see section 4.2.2 for
instructions.
Defining Over/Under Limits
Press the button corresponding to the
Limit Setup.
Note: See section 4.3.1 for information on how to set the
Over/Under limits.
1.800.561.8187
icon. If desired, edit the value
icon to save the changes and
and select Check Mode to Check Counting. Place objects on the pan. The
The CHECK Home screen
Main Display Line
Reference Fields
Functions
button for button icon explanation.
icon to enter