Parts Counting – Check
4.2.3
This feature permits establishing a set quantity of pieces as criteria for similar items that can be quickly checked
against a sample.
The default (or last) Average Piece Weight (APW) is
displayed. Place objects on the pan to display the
Check Status (number of pieces).
4.2.3.1 Edit Settings: To use the Parts Counting-Check mode,
the Over Limit and Under Limit values must be set.
The Average Piece Weight and standard sample size can also
be set. Follow these steps:
Touch the Edit Settings button. The Settings screen appears.
The buttons for APW, Samples, Sample Name, Sample ID,
Batch ID, Clear All.
Note: please refer to section 4.1.1 for details about setting
Sample Name, Sample ID, Batch ID and Clear All.
Over Limit and Under Limit show default (or previously set)
values. You can also touch Batch ID to set an ID for your
samples. To change any value, touch the button. A keyboard
screen appears.
(See keyboard screens, next page.)
To set a new average piece weight, press the
Recalculate APW button.
A new screen appears with the message:
Place reference weight on the pan. Press Accept to
continue.
After following these instructions, press Return to Application
on the main Edit Settings screen.
1.800.561.8187
The PARTS COUNTING – Check Home screen
www.
Press the
Setup
(wrench)
button to enter
the Counting
Setup menu.
Change
Counting
Mode to
Check
Counting.
Main Display Line
Second Display Line
Reference Fields
Application Buttons
Functions
.com
EXPLORER BALANCES
information@itm.com

Advertisement

Table of Contents
loading

Table of Contents