Lumel N30O User Manual

Lumel N30O User Manual

Digital panel meter
Hide thumbs Also See for N30O:

Advertisement

Quick Links

DIGITAL
PANEL METER
N30O TYPE
USER'S MANUAL


Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the N30O and is the answer not in the manual?

Questions and answers

Summary of Contents for Lumel N30O

  • Page 1 DIGITAL PANEL METER N30O TYPE USER’S MANUAL...
  • Page 3 Contents 1. APPLICATION AND METER DESIGN ........5 2. METER SET ..................6 3. BASIC REQUIREMENTS, OPERATIONAL SAFETY ....7 4. INSTALLATION ................8 5. SERVICE ..................12 6. RS-485 INTERFACE ..............36 7. SOFTWARE UPDATING ............50 8. ERROR CODES .................52 9. TECHNICAL DATA ..............53 10. PROGRAMMING ExAMPLES ...........56 11.
  • Page 5: Application And Meter Design

    1. APPLICATION AND METER DESIGN The programmable digital panel meter N30O is destined for measu- rements: number of pulses, frequency, period, worktime, encoder position. Additionally, the meter enables the indication of the current time. The readout field is a LED display, which allows the exposition in colours: red, green and orange.
  • Page 6: Meter Set

    Fig.  View of the N0O Digital Meter 2. METER SET The set is composed of: - N30O meter ......... 1 pc - user’s manual ........1 pc - guarantee card ........1 pc - set of clamps to fix in the panel ..4 pcs - seal ............
  • Page 7: Basic Requirements, Operational Safety

    3. BASIC REQUIREMENTS, OPERATIONAL SAFETY In the safety service scope, the N30O meter meets the requirements of the EN 61010-1 standard. Mentioned below applied symbols mean: - especially important, one must acquaint with this information before connecting the meter. The non-observance of notices marked by this symbol can occasion injures of the personnel and a damage of the instrument.
  • Page 8: Installation

    4. INSTALLATION The meter has separable strips with screw terminals, what enables the connection of external wires of 1.5 mm cross-section for input signals and 2.5 mm for other signals. ´ 45 One must prepare a hole of 92 +0,6 +0,6 mm in the panel, which the thickness should not exceed 6 mm.
  • Page 9 4.1. Lead-out of Signals Signals led out on the meter connectors are presented on the fig. 4. All input signals are separated between them and separated from other circuits. Circuits of successive groups of signals are separated between them. Fig. . Description of Signals on Connection Strips Clamps Description Clamps...
  • Page 10 4.2. Examples of Connections An example of a N30O meter and a inductive sensor connection with an output of NPN and PNP type is presented on the fig. 5. The way to connect a transducer with an output of contactron/relay type is pre- sented on the fig.
  • Page 11 Fig. . Connection of the sensor with the OC Output: a) NPN Type b) PNP Type. Fig. . Connection of the sensor with the Reed Relay/ Relay Output Type...
  • Page 12: Service

    Fig. . Description of the Meter Frontal Plate 5.2. Messages after Switching the Supply on After switching the supply on, the meter displays the meter name N30O and next, the program version in the shape „ x.xx” – where x.xx is the number of the current program version or the number of a custom- made option.
  • Page 13 Þ moving through the menu – choice of level, Þ entry in the mode changing the parameter value, Þ acceptation of the changed parameter value, Þ stop the measurement – when holding down the button, the result on the display is not updated. The measurement is still carried out.
  • Page 14 sing of the minimal value. The pressure of the button combination causes the era- sing of the maximal value. The pressure of the button combination causes the display of the second counter contents. A longer holding down (longer than 3 seconds) causes the reset of the main counter (if the ser- vice of counters from the keyboard is switched on).
  • Page 18 the entry in the matrix, the write of an incorrect code causes the display of the ErCod symbol. The matrix of transitions into the programming mode is presented on the fig. 9. The selection of the level is made by means of the button, however the entry and moving through the parameters of the chosen level is carried out by means of...
  • Page 19 2) setting of the decimal point position (00000., 0000.0, 000.00, 00.000, 0.0000); the button shifts the decimal point to the left, howe- ver the button shifts the decimal point to the right; The pressure of the button during the change of the parameter value will cause the resignation of the write.
  • Page 20 Table 1 Minimal duration time of the high level pulse on the main input. The introduc- tion of a value lower than 0.25 or 5 causes the switching of the length con- trol function of the high signal level off. t_H1 0...60000 The value is given in milliseconds.
  • Page 21 oFF – lack of mathematic operation Mathematic function. On the mea- sqr – ( sured value is carry out additionaly a measured value √ FUnCt chosen mathematic operation. Then sqrt – measured value this value is rescaling by individual Inv – 1/ measured value characteristic.
  • Page 22 On – Control inputs steers the work of the auxiliary counter. Permit external functions: E_In2 Off – Control inputs do not start/stop, erasing influence the auxiliary counter work. The counter is automatically erased Auto2 after reaching this value. The write of -19999...99999 the value 0 switches the function off.
  • Page 23: Parameter Description

    Table 4 dISP Parameter Description Range of changes symbol Minimal position of the decimal point 0.0000 – when displaying the measured value 00.000 – - display format. This parameter is not 000.00 – taken into consideration during the 0000.0 – modes CountH and HoUr.
  • Page 24 Table 5 ALr1, ALr2, ALr3, ALr4 Parameter Description Range of changes symbol P_A1 InP1 – Main input (indicated P_A2 value). Input quantity, steering the alarm. P_A3 InP2 – input of the auxillary counter. P_A4 PrL1 PrL2 Alarm lower threshold. -19999...99999 PrL3 PrL4 PrH1...
  • Page 25 Table 5 dLY1 Delay of alarm switching (time of dLY2 delay by alarm switching on and 0...900 dLY3 switching off. dLY4 Support of alarm signaling. In the situation when the support function is switched on, after the alarm state retrea, the signaling diode is not blanked.
  • Page 26: Rs-485 Interface

    Table 6 Parameter Description Range of changes symbol InP1 – main input (indicated value). Input quantity, on which the P_An analog output has to react. InP2 – input of the auxillary counter. Lower threshold of the analog output. One must give the value, on which we want to obtain the -19999...99999 minimal value of signal on the...
  • Page 27 Table 7 Parameter Description Range of changes symbol Write in manufacturer’s settings. The setup of the value YES causes no – do nothing. the write of standard parameters YeS – causes the write of into the meter. manufacturer’s settings. Values of manufacturer’s param- eters are presented in the table 9.
  • Page 30 Cons1 matematyczne mierzona indywidualna 5.4.4 Individual Characteristic analogowymi N30O meters can recount the measured value into any value thanks Sterowanie to the implemented individual characteristic function. The individual alarmami characteristic causes rescale of the measured value for the expected Interfejs indication range vaule.
  • Page 31 Fig. . Individual Characteristic During the function approximation, one must remember that for the ap- proximation of functions strongly differing from the linear characteristic, higher the number of linearizing segments, smaller the error related to the linearization. If measured values are smallest from H1 then, recalculations will be made on the base of the first straight line calculated on the base of po- ints (H1, Y1) and (H2, Y2).
  • Page 32 5.4.5. Alarm Types The N30O meter is equipped with 2 alarm outputs with NOC contact (make contact) and two alarm outputs with NOC/NCC contact (make and break contact) (option). Each of alarms can work in one of the six modes. The work of alarms in modes is presented in the fig. 12: n-on, n- off, on, off.
  • Page 33 5.4.6 Display Format The N30O meter adapts automatically the display format (precision) to the value of measured quantity. So that the function could be fully used, one must choose the format 0.0000, then the meter will display the me- asured value with the possible highest accuracy.
  • Page 34 5.5. Manufacturer’s Parameters Standard settings of the N30O meter are presented in the table 9. These settings can be restored by means of the meter menu through the choice of the option Set from the menu Ser. Table 9 Parameter symbol...
  • Page 35 00000 CoLdo grEEn CoLbE orAng CoLuP CoLLo 5000 CoLHi 8000 ovrLo -19999 ovrHi 99999 P_A1, P_A2, P_A3, 5, 6, 7, 8 InP1 P_A4 tYP1, tYP2, tYP3, 5, 6, 7, 8 h-off tYP4, PrL1, PrL2, PrL3, 5, 6, 7, 8 1000 PrL4 PrH1, PrH2, PrH3, 5, 6, 7, 8...
  • Page 36 6. INTERFACE RS-485 N30O programmable digital meters have serial interface in RS-485 standards for the communication in computer systems and with other devices fulfilling Master function. An asynchronous com- munication character protocol MODBUS has been implemented. The transmission protocol describes ways of information between devices through the serial interface.
  • Page 37 6.2. Description of the MODBUS Protocol Implementation The implemented protocol is in accordance with the PI-MBUS-300 Rev G of Modicon Company specification. Set of the serial protocol parameters of N30O meters in MODBUS protocol: meter address 1...247, baud rate 4800, 9600, 19200, 38400,...
  • Page 38 16 – Write of n-registers. 17 – Identification of the slave device. 6.4 Register Map The register map of the N30O meter is presented below Notice: All given addresses are physical addresses. In some computer programs, logic addressing is applied, then addresses must be increased of 1.
  • Page 39 6.5. Registers for Write and Readout Table 11 Symbol Range Description 4000 0...7 Input type tYP1 Value Pulse counter Frequency (f < 10 kHz) Frequency (f > 10 kHz) Rotational speed Period Long period Worktime counter Current time Incremental encoder Selection of input quantity recalibration Value Description...
  • Page 40 Selection of input quantity recalibration Value Description 4004 0, 1 SCAL2 Multiplication by the constant Division by the constant Permit for external functions Value Description 4005 0, 1 E_In2 External functions does not influence the auxiliary counter work. External functions control the auxiliary counter work.
  • Page 41 000.00 0000.0 00000 Display colour when the displayed value is lover than coLLo Value Description 4010 0...2 CoLdo green orange Display colour when the displayed value is higher than CoLLo and lower than CoLHi Value Description 4011 CoLbE 0...2 green orange Display colour when the displayed value is higher than CoLHi...
  • Page 42 4015 0...900 dLY1 Delay of alarm 1 switch on and off (in seconds) 4016 0...1 Support of alarm 1 signaling LEd1 Value Description Suport turned off Suport turned on 4017 0, 1 P_a2 Input quantity controlling the alarm Value Description Main input Auxiliary input 4018...
  • Page 43 4023 0...900 dLY3 Delay of alarm 3 switch on and switch off (in seconds) 4024 0...1 Support of alarm 3 signaling LEd3 Value Description Support switched off Support switched on 4025 0, 1 P_a4 Input quantity controlling the alarm Value Description Main input Auxiliary input...
  • Page 44 19200 bit/s 38400 bit/s 57600 bit/s 115200 bit/s 4032 0...3 Transmission mode prot Value Description RTU 8N2 RTU 8E1 RTU 8O1 RTU 8N1 Device address. The write of the value 0 causes the 4033 0...247 Addr interface switching off. Update transmission parameters. Causes the 4034 0...1 sAvE...
  • Page 45 Bit 14 RTC clock. Loss of settings Bit 13 Not used Bit 12 Lack of communication with data memory Bit 11 Wrong settings Bit 10 Manufacturer’ s settings restored Bit 9 Lack of measured values in data memory Bit 8 Reset of the auxiliary counter Bit 7 Output plate is detected...
  • Page 46 Table 12 write Sym- Range Description rea- dout Lower threshold of the display color 7200 7600 CoLLo -19999...99999 change Upper threshold of the display color 7202 7601 CoLHI -19999...99999 change 7204 7602 ovrLo -19999...99999 Lower threshold of the display narrowing 7206 7603 ovrHI...
  • Page 47 Automatic reset of the main counter 7234 7617 Auto1 -19999...99999 Measurement time of the low frequency, speed and periods (table 8a) Constant recalibrating the input quantity 7236 7618 Cons2 -19999...99999 on the auxiliary input Minimal duration of the low pulse level on 7238 7619 0...60000...
  • Page 48 Table 12 Point of the individual characteristic. 7284 7642 -19999...99999 Point No. 11 7286 7643 -19999...99999 Expected value for the point No. 11 Point of the individual characteristic. 7288 7644 -19999...99999 Point No. 12 7290 7645 w/r -19999...99999 Expected value for the point No. 12 Point of the individual characteristic.
  • Page 49 Name of the quantity /readout Constant identifying the device. 7000 7500 Identifier — The value 181 means the N30O meter Status is register describing the 7002 7501 Status — current state of the meter It is a register defining the steering...
  • Page 50: Software Updating

    After starting LPCon’s software COM port, baudrate, transmission mode and adress should be set. It can be done in Options. Then, N30O meter should be selected from Device. Push icon Load to read and save current settings. Open window Lumel Updater (LU) – figure 14b...
  • Page 51 te progress is shown in Messages section. Text Port opened appear after correctly opened port. Putting meter in update’s mode can be done in two ways: remote from LU (with settings from LPCon – port, baudrate, transmission mode and adress) or by turning power on while button pressed.
  • Page 52: Error Codes

    8. ERROR CODES After switching the meter on or during the work, messages about errors can appear. Messages about errors and their reasons are presented below. Table 14 Error message Description Overflow of upper value of the measuring range value or the programmed indication range Overflow of lower value of the measuring range value or the programmed indication range...
  • Page 53: Technical Data

    9. TECHNICAL DATA Measuring Ranges. Table 15 Kind of inputs Indication range Class Number of pulses -19 999..99 999 ±1 pulse Cntr1, Cntr2 Frequency <10 kHz 0.05...99 999 Hz 0.01 1...99 999 kHz Frequency >10 kHz (measuring range up to 0.01 1 MHz) Rotational speed...
  • Page 54 24 V d.c./30 mA Output of auxiliary supply output of OC type, passive npn, Alarm output OC (option) 30 V d.c./30 mA. Input signals voltage 5...36 V d.c., galvanically separated Duration of control signals higher than 10 ms Serial interface RS 485 address: 1...247 mode: 8N2, 8E1, 8O1, 8N1 baud rate: 4.8, 9.6, 19.2, 38.4, 57.6,...
  • Page 55 Additional errors: - from temperature changes: for analog inputs and outputs 50% of the class/10 K Standards fulfilled by the meter: Electromagnetic compatibility: noise immunity acc. to EN 61000-6-2 noise emissions acc. to EN 61000-6-4 Safety requirements: acc. to EN61010-1 standard: isolation between circuits: basic, installation category: III, pollution level: 2,...
  • Page 56: Programming Examples

    10. PROGRAMMING EXAMPLES Example 1. Programming the meter to work with the flowmeter with reed relay output of the following parameters: - pulse constant 1K - 4,3956 dm /pulse = 0,0043956 m /pulse; - minimal flow Q – 0,02 m The parameters of input 1 should be set as follows: - tYP1 should be set to FreqL (frequency for (f<10kHz));...
  • Page 57 - Auto1 – the maximum time interval between pulses for minimum flow - 16 [s]; - FUnCt – mathematical functions - for frequency measurement should be set to oFF; - Cnt1 – set to value 1 (averaging current measurements every second). Then set the individual characteristics: - IndCp –...
  • Page 58: Order Codes

    - after agreeing with the manufacturer. Order example: The code N30O - 1 0 01 00 E 0 means: N30O - programmable N30O digital panel meter 1 - supply: 85...253 V a.c./d.c.
  • Page 59 Code of the backlighted unit Table 17 Code Unit Code Unit Lack of unit kvar Mvar obr/min r.p.m. kvarh mm/min Mvarh m/min kVAh l/min MVAh /min szt/h km/h kg/h on order 1) - after agreeing with the manufacturer.
  • Page 60: Maintenance And Guarantee

    12. MAINTENANCE AND GUARANTEE The N30O digital panel meter does not require any periodical mainte- nance. In case of some incorrect operations: 1. From the Shipping Date, During the Period Given in the Annexed Guarantee Card One should take the meter down from the installation and return it to the Manufacturer’s...
  • Page 62 N30O-09C Lubuskie Zak³ady Aparatów Elektrycznych LUMEL S.A. ul. Sulechowska 1 65-022 Zielona Góra - Poland Tel.: (48-68) 329 51 00 (exchange) Export Department: Fax: (48-68) 329 51 01 Tel.: (48-68) 329 53 02 or 53 04 e-mail: lumel@lumel.com.pl Fax: (48-68) 325 40 91 http://www.lumel.com.pl...

Table of Contents