Page 1
CUSTOMER SERVICE MANUAL NO. 246 According to DIN 58345 Certified from TÜV SÜD Management Service GmbH in accordance to ISO 9001: 2000 Philipp Kirsch GmbH ● Headquarters: 77652 Offenburg ● Okenstraße 103 Phone.: 0049-781-9227-0 ● Fax: 0049-781-9227-200 ● info@kirsch-offenburg.de ● www.kirsch-offenburg.de 1/12...
Page 2
Technical data: Cooler: VKS 2116 K Voltage: 220-240 V 50/60 Hz Cooling capacity: 150 W at tm 0°C Total current consumption: 1.4 A L´Unite compressor: THB4419YWZ TH751WZ Current consumption cross-flow fan: 0.15 A Refrigerant: R 134 a, 80 g (for exact filling quantity see details on the cooler) Battery for the mains-independent acoustic warning system: 6 V 4Ah General technical description of the refrigerator: The refrigerator is equipped with an electronic control unit and an electronic monitoring unit.
Page 3
After being switched on the temperature alarm is not set until the reference value has been reached or after 2 hours. This gives the cooling system the opportunity to reach the working temperature range without the alarm being triggered. If the temperature falls below or rises above the respective set temperature thresholds, an acoustic signal is sounded in the operating unit, the remote warning contact is triggered and the current temperature flashes alternately with the fault message in the display 4 (see fault and status messages).
Page 4
A1 Replacing the display or the control and monitoring unit A2 Both parts are placed below the top plate. Remove the cover plate at the rear side of pull out the work surface. The parts then become visible. Replacing the cross-flow fan: Screw the upper angled part of the air baffle off and the attachment plate with the cross-flow fan off the rear of the interior space and take it out.
Page 5
Fault and status messages (extract from the operating manual) Fault messages are displayed alternating with the current temperature. If several faults have occurred, these are displayed in succession. Display Cause Effect Measure Key switch in position 0 Appliance connected to the mains Switch the appliance on, turn key switch to supply, regulator on stand-by position 1...
Page 6
The value is automatically reset to the original value if no key is pressed for 45 seconds or by simultaneously pressing buttons 3 and 8 for approx. 4 seconds. Parameter Functional description Setting range MED-288 Defrosting interval 0: no automatic defrosting 1 ... 99 hours 12 hours Stop at defrosting temperature 0 ...
Page 7
The value is automatically reset to the original value if no key is pressed for 45 seconds or by simultaneously pressing buttons 3 and 8 for approx. 4 seconds. Parameter Functional description Setting range MED-288 Reference value upon delivery C7 ... C6 5.0 °C Switching sense...
Page 8
The value is automatically reset to the original value if no key is pressed for 45 seconds or by simultaneously pressing buttons 3 and 8 for approx. 4 seconds. Parameter Functional description Setting range MED-288 Upper warning limit A2/A7 ... A6 8.0 K Lower warning limit A7 ...
114407 Key for key switch 216104 Drip dish 530 x 38 x 30 mm 202067 Akku 6 V 4 Ah 222156 Drawer MED-288 226617 Hoop for battery 222158 Aluminium length divider for drawer 330 mm 222159 Cross dievicer H + H 160 mm...