Table 1–3: Warranted Electrical Characteristics - Tektronix P6015A Instruction Manual

1000x high voltage probe
Table of Contents

Advertisement

Specifications
Table 1–3: Warranted Electrical Characteristics
Characteristic
Maximum input voltage
1
DC + peak AC
Peak pulse
Bandwidth (–3 dB)
10-ft cable
25-ft cable
2
Rise Time
10-ft cable
25-ft cable
DC attenuation
1
Characteristic not checked in manual
2
T
(ns) = .35/BW (MHz)
r
1–24
Information
1.5 kV to 20 kV. See frequency derating curve in
Figure 1–4. (DC plus peak AC rating is limited to
temperatures below 35° C.)
a
40 kV
(Never exceed 20 kV rms)
Duty cycle derating – 100 ms maximum duration at 10%
maximum duty cycle. See duration and duty cycle
derating curve in Figure 1–5.
Altitude derating – Peak pulse derated linearly from
40 kV at 8000 feet (2440 m) to 30 kV at 15,000 feet
(4570 m) altitude.
Relative Humidity (RH) derating – Voltage derated with
increasing temperature and relative humidity (see
Figure 1–7).
Test conditions: Test oscilloscope bandwidth must be
≥100 MHz, Z
= 25 W
source
75 MHz
25 MHz
≤4.67 ns (calculated from bandwidth)
≤14 ns (calculated from bandwidth)
1000:1 ±3% (Excluding
oscilloscope error)
P6015A Instruction Manual
Test conditions: Oscillo-
scope input resistance
must be 1 MW ±2%