Warranted Characteristics
Table 2–7: Warranted Characteristics — Signal Acquisition System (Cont.)
Name
Analog Bandwidth, DC-50
DC-1 M Coupled
Cross Talk (Channel Isolation)
Input Impedance, DC-1 M Coupled
Input Impedance, DC-50
Input Voltage, Maximum, DC-1 M ,
AC-1 M , or GND Coupled
Input Voltage, Maximum, DC-50
AC-50
Coupled
Lower Frequency Limit, AC Coupled
Delay Between Channels, Full Band-
width, Equivalent Time
1
Net Offset = Offset – (Position
dynamic range. Offset Accuracy is the accuracy of this Voltage level.
2
The samples must be acquired under the same setup and ambient conditions.
3
To ensure the most accurate measurements possible, run an SPC calibration first. When using the oscilloscope at a
Volts/Div setting
5 mV/div, an SPC calibration should be run once per week to ensure that instrument performance
levels meet specifications.
4
The limits given are for the ambient temperature range of 0 C to +30 C. Reduce the upper bandwidth frequencies by
4.0 MHz for each C above +30 C.
5
The AC Coupled Lower Frequency Limits are reduced by a factor of 10 when 10X, passive probes are used.
Table 2–8: Warranted Characteristics — Time Base System
Name
Accuracy, Long Term Sample Rate and
Delay Time
2–10
Description
Coupled or
Volts/Div
5 mV/div – 10 V/div
2 mV/div – 4.98 mV/div
1 mV/div – 1.99 mV/div
100:1 at 100 MHz and
equal volts/division settings
1 M
0.5% in parallel with 10 pF 2 pF
Coupled
50
1% with VSWR
300 V CAT II; derate at 20 dB/decade above 1 MHz
or
5 V
, with peaks
RMS
10 Hz when AC–1 M Coupled;
250 ps for any two channels with equal volts/division and coupling settings
Volts/Div). Net Offset is the nominal voltage level at the center of the A-D converter
Description
25 ppm over any
4
Bandwidth
DC – 500 MHz
DC – 350 MHz
DC – 250 MHz
30:1 at the derated bandwidth for any two channels having
1.3:1 from DC – 500 MHz
30 V
200 kHz when AC – 50
1 ms interval
TDS 510A Performance Verification and Specifications
5
Coupled