Specifications - AEMC J93 User Manual

Ac/dc current probe
Hide thumbs Also See for J93:
Table of Contents

Advertisement

4.1 Reference Conditions
Quantity of influence
Temperature
Relative humidity
DC signal
AC signal
Supply voltage
External electric field
External DC magnetic field
(earth's field)
External AC magnetic field
Position of the conductor
4.2 Electrical
Measurement Range
Sensitivity
Output impedance
Frequency
Primary
Current
(AC/DC)
Amplitude
error
± 2% ± 2.5A ± 1.5% ± 2.5A
(see curve
in §4.4.1)
Phase error
(see curve
in §4.4.2)
Zero adjustment: ± 200A in 25 turns
AC/DC Current Probe Model J93
Shop for AEMC products online at:

4. SPECIFICATIONS

With an AC signal of which the distortion factor is <0.1%
50 to 100
100 to 500
ShopAEMC
www.
Reference Values
23°C ±5°C
20 to 75% RH
45 to 65Hz with a DC signal <0.1%
J93: 9V ± 0.1V
Zero
< 40 A/m
Zero
centered in the probe jaws
50A to 3500A
50A to 5000A
1V / 3500A
≥ 100kW
DC to 3kHz (-3 dB typical)
2000 to 3500 3500 to 5000
500 to 2000
± 1%
.com
ac
dc
(DC only)
± 1%
± 1%
1.5°
1.877.766.5412