Ph Calibration; Why Calibrate; When Do You Have To Calibrate; Carrying Out Automatic Calibration (Autocal) - wtw Multi 9310 Operating Manual

Digital ph / orp / d.o. / conductivity meter for digital ids sensors
Table of Contents

Advertisement

pH value
30
mode:
Temperature
sensor
yes
-
5.2

pH calibration

5.2.1 Why calibrate?

pH electrodes age. This changes the zero point (asymmetry) and slope of the
pH electrode. As a result, an inexact measured value is displayed. Calibration
determines and stores the current values of the zero point and slope of the
electrode.
Thus, you should calibrate at regular intervals.

5.2.2 When do you have to calibrate?

Routinely within the framework of the company quality assurance
When the calibration interval has expired

5.2.3 Carrying out automatic calibration (AutoCal)

Make sure that in the sensor menu, Buffer menu, the buffer set is correctly
selected (see section 9.1.1 S
Use one to five buffer solutions of the selected buffer set in any order.
Below, calibration with Technical buffers (TEC) is described. When other buffer
sets are used, other nominal buffer values are displayed. Apart from that, the
procedure is identical.
If single-point calibration was set in the menu, the calibration proce-
dure is automatically finished with the measurement of buffer solu-
tion 1 and the calibration record is displayed.
1.
Connect the pH sensor to the meter.
The pH measuring window is displayed.
2.
Keep the buffer solutions ready.
3.
Start the calibration with <CAL>.
The calibration display for the first buffer appears (voltage display).
Resolution of the
temp. display
0.1 °C
1 °C
ETTINGS FOR P
Temp. measurement
Automatic with temperature
sensor
Manual
H
, page 62).
MEASUREMENTS
ba75937e03
Multi 9310
12/2011

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Multi 9310 and is the answer not in the manual?

Subscribe to Our Youtube Channel

This manual is also suitable for:

Multi 9310p

Table of Contents