Checking The Functions Before Using The Instrument; Battery; Temperature; Ullage - Honeywell HERMetic UTImeter Gtex Chem Operation And Service Manual

Portable gas tight electronic gauging device ullage - temperature - interface detector
Table of Contents

Advertisement

Checking the functions before using the
9.8
instrument
Before installing the HERMetic instrument as described in
section 9.9, the following tests are recommended to ensure
that the instrument is ready to work.

9.8.1 Battery

Refer to section 10.2 "Checking the battery".

9.8.2 Temperature

Switch on the unit.
The buzzer shall beep every 2 sec.
When the temperature is displayed, check that it shows the
surrounding temperature.
position A

9.8.3 Ullage

Switch on the unit.
The buzzer shall beep every 2 sec.
Check the ullage in a glass of water.
Check the ullage by immersing the ultrasonic gap sensor
but not the electrodes (position A); The buzzer shall beep
continuously.

9.8.4 Interface

Switch on the unit.
The buzzer shall beep every 2 sec.
Check the interface in a glass of water.
Check the interface by immersing the interface electrodes also
(position B). The buzzer shall beep intermittently.
Figure 9-7
position B
33